Specification | 1551A Ex 1551A-9 1551A-12 1551A-20 |
1552A Ex 1552A-12 1552A-12-DL |
Temperature range | -50 °C to 160 °C (-58 °F to 320 °F) |
-80 °C to 300 °C ( -112 °F to 572 °F) |
Accuracy (1 year) | ± 0.05 °C (± 0.09 °F) | |
Display units | °C, °F | |
Sensor type | 100 Ω thin-film RTD | 100 Ω wire-wound PRT |
Probe temperature coefficient | 0.00385 Ω/Ω/°C nominal | |
Sensor length | ≤ 10 mm | ≤ 30 mm |
Sensor position (from sheath tip) | 3 mm (0.1 in) | |
Minimum immersion depth 1 | 7 cm (2.8 in) | 12 cm (4.8 in) |
Probe sheath material | SST | |
Response time | 4.8 mm diameter probe (3/16 in): 14 seconds 6.35 mm diameter probe (1/4 in): 21 seconds |
|
Probe hysteresis | ± 0.01 °C | |
Temperature resolution | Selectable: 0.1, 0.01, 0.001 (default 0.01) | |
Sample rate | Selectable: 0.5 seconds, 1 seconds, 2 seconds (default 1 seconds) | |
Operating temperature range readout | -10 °C to 50 °C (14 °F to 122 °F) | |
Storage temperature | -20 °C to 60 °C (-4 °F to 140 °F) 0 % to 95 % RH, non-condensing |
|
Optional data logging 2 | Up to 10,000 time-stamped readings stored to internal memory | |
Logging intervals 2 | 2 seconds, 5 seconds, 10 seconds, 30 seconds, or 60 seconds; 2 minutes, 5 minutes, 10 minutes, 30 minutes, or 60 minutes |
|
Damping | Moving average of the most recent 2 to 10 readings (ON/OFF, 2, 5, 10) |
|
Communications | RS-232 stereo jack (access calibration parameters only) | |
AC power | None | |
DC power | Three AAA batteries, typical battery life of 300 hours without LCD backlight |
|
EMC compliance | EN61326:2006 Annex C; CISPR II Edition 5.0-2009; Class “B” |
|
Enclosure rating | IP50 | |
Electronics dimensions (H x W x D) | 114 mm x 57 mm x 25 mm (4.5 in x 2.25 in x 1.0 in) |
|
Weight | 196 g (6.9 oz) | |
Calibration (included) | NVLAP-accredited, NIST-traceable | |
Characterization | CVD | ITS-90 |