Ground Bond Test Mode |
Output Voltage (Open Circuit Voltage) |
Range: Resolution: Accuracy: |
3.00 – 8.00 VAC/DC 0.01 VAC/DC ± (3% of setting + 3 counts) |
Output Frequency |
50 or 60 Hz, User Selectable/DC |
Output Current |
Range:
Resolution: Accuracy: |
0 – 150 mΩ for 30.01 – 40.00 A 0 – 200 mΩ for 10.01 – 30.00 A 0 – 600 mΩ for 1.00 – 10.01 A 0.1 A ± (3% of setting + 3 counts |
Maximum Loading |
Range:
Resolution: Accuracy: |
1.00 – 10.00 A, 0 – 600 mΩ 10.01 – 30.00 A, 0 – 200 mΩ 30.01 – 40.00 A, 0 – 150 mΩ 1 mΩ ± (2% of setting + 2 counts) |
HI and LO-Limit Resistance |
Range:
Resolution: Accuracy: |
0 – 150 mΩ for 30.01 – 40.00 A 0 – 200 mΩ for 10.01 – 30.00 A 0 – 600 mΩ for 1.00 – 10.01 A 1 mΩ ± (2% of setting + 2 counts |
HI and LO-Limit Voltage |
Range: Resolution: Accuracy: |
0.00 – 6.00 V 0.01 ± (2% of settings + 2 counts |
Dwell Time Setting |
Range: |
0, 0.5 – 999.9 sec (0=Continuous |
Ω Offset Capability |
Range: Resolution: Accuracy: |
0 – 100 mΩ 1 mΩ ± (2% of setting + 2 counts) |
V Offset Capability |
Range: Resolution: Accuracy: |
0.00 – 4.00 V 0.01 V ± (2% of setting + 2 counts) |
Current Display |
Range: Resolution: Accuracy: |
0.00 – 40.00 AAC/DC 0.01 AC/DC ± (3% of reading + 1 count) |
Voltage Display |
Range: Resolution: Accuracy: |
0.00 – 8.00 VAC/DC 0.01 AC/DC ± (2% of reading + 2 counts) |
Ohmmeter Display |
Range: Resolution: Accuracy: |
0 – 600 mΩ for 1.00 – 5.99 A 1 mΩ ± (3% of reading + 3 counts) |
|
Range: Resolution: Accuracy: |
0 – 600 mΩ for 6 – 40 A 1 mΩ ± (2% of reading + 2 counts) |